Traceability- Hardware and Mechanical Requirements

Hi there,

Newbie tester here, currently writing verification test protocols based on client's requirement documents.

Some problematic requirements I encountered:

"The enclosure shall include a board."

"Distance between each tooth shall not exceed 1 mm."

Does such requirements need to actually verified in the test cases? If so, can I cover it by a "hardware inspection" test cases?

Some clarifications:

I'm the only tester and no one at work can answer me.

The client will eat me alive if I'll ask him how to do my job (understandably)

The device is medical, I don't know if they submit it to the FDA but one way or the other- the documents are expected to be very high standard.

Thanks in advance!